Digital Control & Safety:
- Parameters can be set and modified when the centrifuge is in standby mode; programs can be locked.
- During operation, all parameters except rotor type can be adjusted and locked.
- Maximum speed limit function: rotor type must match the actual rotor used. If an incorrect maximum speed higher than the rotor’s rated value is set, the system will issue an error warning.
Speed Accuracy:
- Within the operating speed range:
- Variable frequency speed control: ±50 r/min.
- Voltage regulation speed control: ±200 r/min.
Noise & Vibration:
- Noise level ≤ 65 dB under maximum speed and full load.
- Whole-machine vibration amplitude ≤ 0.1 mm.
Timing & Operation:
- Timer range ≥ 59 minutes.
- Continuous working time ≥ 4 hours without abnormal operation.
Temperature Control:
- Air-cooled chamber; chamber temperature ≤ ambient temperature + 10 °C during operation.
Display Functions:
- Relative centrifugal force (RCF) can be converted and displayed in real time during operation.